Sherescan specifications: |
Mains supply | | 230/115 V AC, 50-60 Hz | | | | Power consumption | | 130 VA | | | | Fuse (20x5 mm) | For 230 V | 1.6 A (slow blow) | | For 115 V | 3.15 A (slow blow) | Measuring environment | Temperature | 15-30 °C | | Humidity level | 80 % | | Vibration free | --- | | | | Interface | | Serial, 9 pin SUB D | | | | Measuring probe | Tip material: Tungsten Carbide Tip radius: 100 μm Tip spacing: 635 μm Pin pressure: 60 - 150 gr/pin (user adjustable) Manufacturer: Jandel | | | | | Overall dimensions | Width | 809 mm | | Height | 350 mm | | Depth | 515 mm | | | | Total weight | | 28 kg | | | | Min. table size | Width (Including monitor) | 800 mm (1200 mm) | | Depth | 600 mm | | | | Sample size | Thickness | 130-1000 µm | | Length (X-axis) | 0-215 mm | | Width (Y-axis) | 0-215 mm | | Sample shape | Square, semi-square, rectangle, circle | Sample material | Multi crystalline and mono crystalline silicon P-type and N-type With and without emitter Textured and non-textured Specific resistivity 10 Ω.cm or less
| | | | | Measurement positions | Free; 1 - 10.000 points; orthogonal | | | | | Measurement uncertainty | 0.5% (electrical) | | | | | Compliance | Up to 55 V | | | | | Measurement range | 10 Ω/sq to 200 Ω/sq | |
|
Sherescan operating software
The Sherescan instrument comes with operating software on a CD. No additional software or hardware is needed. To get an idea of functionality and appearance of software, you can take a look at the panels used in the software. It is also possible to download the software and manual. An example data file to read with the software can be downloaded here. |
Required computer specifications:
Processor | Pentium 1 GHz minimum | Internal memory | 256 MB or more | CD drive | | Hard disk | 50 MB minimum | Keyboard | | Mouse | | Communication | Serial port COM1 or COM2, baud rate 57600 bps. USB also possible with USB to serial converter. | | | Operating system | Windows 98 or newer | | | Display resolution | 1024 x 768 pixels minimum |
|